• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1804400-23
Medical Device Problem Code Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.The additional nc trek neo device referenced in b5 is filed under separate medwatch report number.D4: the udi is unknown due to the part/lot number was not provided.
 
Event or Problem Description
It was reported that procedure was to treat a 50% stenosed de novo lesion in the in the proximal left anterior descending (plad) artery with heavy calcification.The 4mm xience stent was implanted.The 4x8mm nc trek neo balloon dilatation catheter (bdc) was used for post dilatation; however, the balloon ruptured at 8 atmospheres (atm).Therefore, the bdc was removed from the patient and the stent was noted to be indented.A non-abbott balloon was used for further post-dilation of the stent indentation.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
 
Event or Problem Description
It was reported that procedure was to treat a 50% stenosed de novo lesion in the in the proximal left anterior descending (plad) artery with heavy calcification.The 4mm xience stent was implanted.The 4x8mm nc trek neo balloon dilatation catheter (bdc) was used for post dilatation; however, the balloon ruptured at 8 atmospheres (atm).Therefore, the bdc was removed from the patient and the stent was noted to be indented.A non-abbott balloon was used for further post-dilation of the stent indentation.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported difficult or delayed activation.The reported treatment appears to be related to operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18904306
Report Number2024168-2024-03310
Device Sequence Number18425794
Product Code NIQ
UDI-Device Identifier08717648233494
UDI-Public08717648233494
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/24/2026
Device Catalogue Number1804400-23
Device Lot Number3060641
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/23/2024
Supplement Date Received by Manufacturer03/19/2024
Initial Report FDA Received Date03/14/2024
Supplement Report FDA Received Date03/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NC TREK NEO RX 4.00 X 8MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-