|
EIT EMERGING IMPLANT TECHNOLOGIES GMBH IMPLANT INSERTER SH CONNECTION; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
|
Back to Search Results |
|
| Catalog Number |
TFT30101 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: b3: unknown event date.E1: initial reporter is j&j company representative h3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in portugal as follows: it was reported the device arrived at the warehouse broken inside the set.No patient or surgical involvement.This report is for one implant inserter sh connection for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: part # tft30101.Lot # ee21di0137.Batch1: lot qty of (b)(4) units were released on 31 mar 2022 with no discrepancies.No ncrs were generated during production.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the implant inserter pin sh of the implant inserter sh connection was broken at the middle of the shaft.The broken fragment is inside of the inserter connection.A dimensional inspection was not performed due to the post-manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.The overall complaint was [confirmed as the observed condition of the implant inserter sh connection would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Corrected data: h4.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|