• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNIER MEDTECH AMERICA, INC. DORNIER HOLMIUM LASER FIBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNIER MEDTECH AMERICA, INC. DORNIER HOLMIUM LASER FIBER Back to Search Results
Model Number HOL270SHE
Medical Device Problem Codes Fracture (1260); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The laser fiber returned by the complainant was evaluated and confirmed to be broken as reported.No manufacturing defects or inconsistencies which could have caused or contributed to the complaint as reported were revealed.It is confirmed that all laser fiber units manufactured by dornier are 100% inspected via visual evaluation as well as power performance testing prior to release for distribution which confirms the operational capacity as well as the state of the device.Dornier laser fibers are fragile, and must be handled with care as indicated on the valid dornier laser fiber ifu.It is likely the root cause of this complaint was related to a mechanical force placed on the laser fibers which caused or contributed to the fiber breakage.
 
Event or Problem Description
A notification was received regarding a holmium fiber unit which was reported to have broken during usage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORNIER HOLMIUM LASER FIBER
Common Device Name
HOLMIUM LASER FIBER
Manufacturer (Section D)
DORNIER MEDTECH AMERICA, INC.
1155 roberts blvd
suite 100
kennesaw GA 30144
Manufacturer (Section G)
DORNIER MEDTECH AMERICA, INC.
1155 roberts blvd
suite 100
kennesaw GA 30144
Manufacturer Contact
amber stallings
1155 roberts blvd
suite 100
kennesaw 30144
7705146204
MDR Report Key18904636
Report Number1037955-2024-00005
Device Sequence Number15433634
Product Code GEX
UDI-Device Identifier00810116020805
UDI-Public01008101160208051728040710F1323S
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K121938
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHOL270SHE
Device Lot NumberF1323S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2024
Initial Report FDA Received Date03/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-