| Model Number |
PCF-H180AL |
| Medical Device Problem Codes |
No Display/Image (1183); Image Display Error/Artifact (1304); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned and the investigation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported, the colonovideoscope image had lines, no image, and an incompatible scope error (e315).The issue occurred during a procedure and resulted in a few minutes delay.The procedure was completed successfully with the same device and there was no report of patient impact and/or harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was evaluated by olympus, and the reported failure (¿image has lines/no image, error e315¿) was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 14 years since the subject device was manufactured.Based on the results of the investigation, error code e315 likely occurred due to fluid invasion of the scope connector.However, the root cause could not be determined.This supplemental report includes a correction to g2 to provide information that was inadvertently not included in the initial medwatch.Olympus will continue to monitor field performance for this device.
|
| |
|