| Model Number |
UNIVERSAL GLENOID - CENTRAL SCREW 20MM |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the prosthesis had to be removed after 5 years as the glenoid sphere became loosened.No further information received.Update dw 11-mar-2024: further information were provided that only the glenosphere became loosened.The plate and screws did not have any issues but were replaced during the revision including the inlay and the glenosphere.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint allegation was confirmed.One unpackaged ar-9165-20 batch 170156713 was received for investigation.Functional testing doesn't apply to this device.The visual inspection revealed that the hex geometry was damaged.The most likely cause of the reported failure is a particular patient event.According to dfu-0189-1 at revision 0.Precautions.7.Deformation of the operative site, which can prevent or impede anchoring of the implant.9.Allergic reactions to implant materials.
|
| |
|