The reported event could be confirmed, since the device was returned, and matches the alleged failure.The received screwdriver, self-holding, and long screwdriver were visually inspected in which device was found to be in good condition.But we see discoloration of the screwdriver blade, on different spots throughout the periphery of the shaft.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.The cleaning, sterilization, inspection and maintenance (l24002000) states: ¿the recommended concentrations and times for device immersion in the cleaning solutions and/ or disinfectants given by the detergent manufacturers shall be observed.If these concentrations and times are exceeded significantly, discoloration or corrosion could occur with some materials.This could also happen if rinsing after cleaning and/or disinfecting is insufficient.¿ [original statements]: no indications of material, manufacturing or design related problems were found during the investigation.Based on the investigation, the root cause was attributed to being user related.If any further information is provided, the complaint report will be updated.
|