| Catalog Number |
1688710105 |
| Medical Device Problem Codes |
No Display/Image (1183); Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.The reported failure mode will be monitored for future reoccurrence.Alleged failure: no video output.Probable root cause: cables; hdmi cables; connectors; digital board; power supply; filter fuse; ac inlet board; main board; transition board; intermittence between transition board and ch connector; software; camera head (sensor, sensor cable).Coupler.Problem toggling light source.Connected monitors.Leaking.Poor grounding.No incorrect communication with light source.Soaking cap disconnection during sterilization.Electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge.Cybersecurity attack pendulum ch inertial measurement unit use error.
|
| |
|