| Model Number |
24674 |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: initial reporter facility name: (b)(6) hospital.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon pinhole occurred.The target lesion was located in the lower limb.An 8.0 x 100, 135cm mustang balloon catheter was advanced for dilation.During preparation, the pressure of the balloon tip was not enough, and it was leaking water.A visible pinhole was noted on the balloon material.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon pinhole occurred.The target lesion was located in the lower limb.An 8.0 x 100, 135cm mustang balloon catheter was advanced for dilation.During preparation, the pressure of the balloon tip was not enough, and it was leaking water.A visible pinhole was noted on the balloon material.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned product consisted of a mustang.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A longitudinal tear was identified in the balloon material.The tear measured 12mm in length and extended from a position 6mm proximal of the distal markerband to 5mm distal of the distal markerband.A visual examination observed damage to the tip of the device.E1: initial reporter facility name: (b)(6) school of medicine.
|
| |
|