EXACTECH, INC. EQUINOXE REVERSE 46MM HUMERAL LINER +0; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-46-00
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Joint Dislocation (2374); Joint Laxity (4526)
Date of Event 02/08/2024
Type of Reportable Event Serious Injury
Event or Problem Description
As reported by the equinoxe shoulder study, the 58-year-old male patient had a right tsa on (b)(6) 2022.The patient presented with disassociation of polyethylene - patient has experienced chronic instability of the shoulder since pod2, then in one subluxation event, the polyethylene dissociated.The outcome of this event is considered resolved and the report indicates that the action taken is revision on (b)(6) 2024.The case report form indicates that this event is definitely related to the device and/or to the procedure.This event report was received through clinical data collection activities and no device return is anticipated.
 
Additional Manufacturer Narrative
(d10) concomitant device(s): 300-30-06 - equinoxe preserve stem 6mm: (b)(6).320-10-00 - equinoxe reverse tray adapter plate tray +0: (b)(6).320-10-46 - glenosphere 46x21mm inset: (b)(6).320-15-04 - rs glenoid plate r post aug, 8 deg, right: (b)(6).320-15-05 - eq rev locking screw: (b)(6).320-20-00 - eq reverse torque defining screw kit: (b)(6).320-20-42 - eq rev compress screw lck cap kit, 4.5 x 42mm: (b)(6).320-20-42 - eq rev compress screw lck cap kit, 4.5 x 42mm: (b)(6).320-20-46 - eq rev compress screw lck cap kit, 4.5 x 46mm: (b)(6).(h3) pending evaluation.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported cannot be confirmed from the information provided but may be the result of disassembly, prosthesis wear, dislocation, and/or instability.However, the reported disassembly, prosthesis wear, dislocation, and instability could not be confirmed.Potential contributions of user and patient-related considerations to the event, an unreported traumatic event, incomplete seating of the humeral liner during implantation, forceful contact between the liner and glenoid bone (scapular notching), or a combination of the above to the reported disassembly of the humeral liner cannot be determined as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE REVERSE 46MM HUMERAL LINER +0
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
miguel sosa
2320 nw 66th ct.
gainesville, FL 32653
3523771140
MDR Report Key18909371
Report Number1038671-2024-00513
Device Sequence Number6971243
Product Code KWT
UDI-Device Identifier10885862086730
UDI-Public10885862086730
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number320-46-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer05/06/2025
Initial Report FDA Received Date03/14/2024
Supplement Report FDA Received Date05/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Patient Age58 YR
Patient SexMale
Patient Weight92 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-