• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RAPTORMITE 3.0 PK W/NDL, 2 #0 ULTRABRA; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RAPTORMITE 3.0 PK W/NDL, 2 #0 ULTRABRA; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72201806
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy procedure, the raptormite implant took out from the implant box insert to patient bone, the anchor suture very soft and bend and difficult to insert repair.It was like broken.The procedure was successfully completed using a back-up device.There was a surgical delay greater than 30 minutes and no further complications were reported.
 
Additional Manufacturer Narrative
The reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The insertion device was returned with the anchor and suture string.The proximal end of the anchor is deformed and it is leaning off the insertion device.There is biological debris on the returned items.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device or an inadvertent impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPTORMITE 3.0 PK W/NDL, 2 #0 ULTRABRA
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18909630
Report Number1219602-2024-00508
Device Sequence Number6971250
Product Code MBI
UDI-Device Identifier03596010619082
UDI-Public03596010619082
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K071586
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72201806
Device Lot Number2101796
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2024
Supplement Date Received by Manufacturer04/19/2024
Initial Report FDA Received Date03/14/2024
Supplement Report FDA Received Date04/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-