| Model Number |
VTICM5_12.1 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/30/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation - lens work order search: one similar complaint type event reported for units within the same lot.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 12.1mm vticm5_12.1 implantable collamer lens of -10.00/3.0/180 (sphere/cylinder/axis) was implanted vertically into the patient's right eye (od) on (b)(6) 2023.On (b)(6) 2023 the lens was explanted due to low vault.A spherical lens of the same length, different power was later implanted horizontally and the problem resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: the lens was returned dry in a microcentrifuge vial with residue/debris on the product.Visual inspection found optic torn, and haptic bent and stretched out.Dimensional inspection found the lens to be within specifications.Claim #: (b)(4).
|
| |
|