H3: other code and h6: code b20 - the device remains implanted in the patient.Therefore, a device evaluation could not be performed.H6: code b14: analysis of production records is underway.Per gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and/or require intervention include, but are not limited to: endoleak.The patient referenced in this event file is enrolled in a post market prospective observational cohort registry designed to obtain data on device performance and clinical outcomes of patients treated with gore endovascular aortic products through the global registry for endovascular aortic treatment (great).Medidata rave® is the clinical study database (csd), capturing the data through the grt 10-11 module.The above information was obtained from the csd.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
On (b)(6) 2016, it was reported that a patient presented with a abdominal aortic aneurysm (measured 53 mm) was treated with a gore® excluder® aaa endoprosthesis.The patient underwent a series of follow-up controls without any significant aneurysm growth.On (b)(6) 2018, the patient's aneurysm was measured 59 mm due to a type ii endoleak and underwent embolization on (b)(6) 2018.A slow but continues growth was still observed and on (b)(6) 2023, the size increased to 76 mm.Embolization was performed again on (b)(6) 2023.No further sac enlargement was detected during the follow-ups (the last one documented on (b)(6) 2024).
|
|
As part of our routine quality procedures, each batch of devices undergoes comprehensive quality control testing and inspections prior to release for distribution.Gore reviewed the manufacturing records associated with the reported lot number and verified that all release criteria had been met.Based on the incident description and the subsequent investigation, no further information was provided to gore, we are unable to determine the cause of this incident and assign a root cause.H6: added medical device problem code a24.Updated investigation findings code to c19 - no device problem found.Code c21 is no longer applicable.Updated investigation conclusions code to d15 and d12.Code d16 is no longer applicable.
|