|
DEPUY MITEK LLC US 5X12 MILAGRO INTERERENCE SCREW; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
231814 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Report 2 of 4 for (b)(4).It was reported by sales rep that during an unspecified surgical procedure on (b)(6) 2024, it was observed that a 8x15mm milagro interference screw, 5x12mm milagro interference screw, 8x23mm milagro advance interference screw and 10x23mm milagro advance interference screw devices were broken.It was reported that the patient had extremely hard bone, and it was a revision case.It was reported that the same large tunnels were used and the doctor wanted to start with a larger screw but it would not drive even after notching and tapping four times.According to the reporter, multiple screws were tried, they all ended up breaking and the broken screw stayed inside the patient as it wasn't badly broken since it was still holding the graft in place.It was reported that the doctor left the left broken screw inside patient as it held the graft but broke in half.There was no delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: concomitant devices: 8x15mm milagro interference screw, 8x23mm milagro advance interference screw, 10x23mm milagro advance interference screw, on (b)(6) 2024.(b)(4) e3: reporter is a j&j sales representative.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The investigation has been updated to reflect the correct information: investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information or the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device lot number (9l04453), and no non-conformance was identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
| |
|
Search Alerts/Recalls
|
|
|