MEDTRONIC IRELAND VALIANT CAPTIVIA - FF; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number VAMF4036C150TU
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Rupture (2208)
Date of Event 03/05/2024
Type of Reportable Event Death
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant captivia stent graft was intended to be implanted during the endovascular treatment of a 88mm thoracoabdominal aneurysm.The patient's aorta was noted as double s-shaped with small access vessels at the iliacs.It was reported that during the index procedure the patient's iliacs were too small to place the valiant captivia so a conduit was placed for access.Once the wires were placed a sentrant sheath was placed into the conduit and the vamf4036c150tu was implanted into the patient.Power injector run was taken to confirm placement before deployment and it was noticed that contrast was stagnant in the aorta.The patient deteriorated and cpr was administered and the graft was removed.The patient expired.Per the physician the cause of death was anatomy related due to unforeseen anatomical complications including the small access requiring a conduit, double s-shaped aorta, and the patients' body habitus was small.The patient had complex and diseased anatomy, and in an attempt to fix the thoracic aneurysm, the an infrarenal aneurysm ruptured.No additional sequelae were reported and the patients is expired.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA - FF
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18912845
Report Number9612164-2024-01293
Device Sequence Number6971346
Product Code MIH
UDI-Device Identifier00643169400177
UDI-Public00643169400177
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/23/2026
Device Model NumberVAMF4036C150TU
Device Catalogue NumberVAMF4036C150TU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2024
Initial Report FDA Received Date03/15/2024
Date Device Manufactured01/24/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Death; Required Intervention;
Patient Age77 YR
Patient SexFemale
-
-