| Model Number |
M00570540 |
| Medical Device Problem Codes |
Positioning Failure (1158); Material Deformation (2976); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a wallflex biliary covered stent was to be used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.During the procedure, the stent did not expand.The stent was removed from the patient using forceps, and the procedure was completed with another wallflex biliary stent.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: imdrf device code a150101 captures the reportable event of stent failure to expand.
|
| |
|
Additional Manufacturer Narrative
|
|
Blocks b5 and h6 were updated based on the additional information received on march 17, 2024.Block h11: a wallflex biliary stent was received for analysis.Visual examination of the returned device found the stent fully covered and undeployed.The stainless-steel handle was bent.Functional inspection was performed by manually gripping the outer sheath and pulling the tip distally to deploy the stent.The stent was expanded correctly.No other problems were noted to the stent and delivery system.According to information received from the complainant on march 17, 2024, confirming that the device was bent, and the stent could not be deployed.This event is no longer considered an mdr reportable event.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a wallflex biliary covered stent was to be used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.During the procedure, the stent did not expand.The stent was removed from the patient using forceps, and the procedure was completed with another wallflex biliary stent.There were no patient complications reported as a result of this event.Additional information received on may 17, 2024: it was reported that the device was bent, and the stent could not be deployed.
|
| |
|