Model Number PM2136
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a follow up in clinic, elective replacement indicator (eri) alert was noted on the device, however, the battery voltage was above eri level.Abbott technical support was contacted, the session records were reviewed, and the eri alert was suspected to be a false alert.The device was explanted and replaced to resolve the event.The patient was in stable condition.
 
Additional Manufacturer Narrative
The reported event of false elective replacement indicator (eri) was confirmed.The pacer was received in normal working conditions with battery voltage above eri.Device triggered eri alert is consistent with autocapture and rate responsive settings turned on.Device image shows autocapture on.When automatic settings are programed, such as autocapture, the device output would adjust itself to accommodate the patient¿s needs for pacing and thus affect the estimated longevity of the device.Electrical and mechanical analysis performed, indicated normal device functionality.The implant duration exceeds the projected longevity based on field settings, battery is still normal and above eri.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUSTAIN XL DR
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18913009
Report Number2017865-2024-35090
Device Sequence Number12452609
Product Code NVZ
UDI-Device Identifier05414734504713
UDI-Public05414734504713
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2014
Device Model NumberPM2136
Device Lot Number3798886
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2024
Supplement Date Received by Manufacturer07/15/2024
Initial Report FDA Received Date03/15/2024
Supplement Report FDA Received Date07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/06/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-