| Model Number |
PM2136 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a follow up in clinic, elective replacement indicator (eri) alert was noted on the device, however, the battery voltage was above eri level.Abbott technical support was contacted, the session records were reviewed, and the eri alert was suspected to be a false alert.The device was explanted and replaced to resolve the event.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of false elective replacement indicator (eri) was confirmed.The pacer was received in normal working conditions with battery voltage above eri.Device triggered eri alert is consistent with autocapture and rate responsive settings turned on.Device image shows autocapture on.When automatic settings are programed, such as autocapture, the device output would adjust itself to accommodate the patient¿s needs for pacing and thus affect the estimated longevity of the device.Electrical and mechanical analysis performed, indicated normal device functionality.The implant duration exceeds the projected longevity based on field settings, battery is still normal and above eri.
|
| |
|