Model Number 24674
Medical Device Problem Codes Entrapment of Device (1212); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3 date of event: event date was not provided at the time of reporting.The first day of the month of the aware date was used.
 
Event or Problem Description
It was reported that a balloon rupture occurred.The target lesion was located in the subclavian artery.A 9.0 x 80, 75cm mustang balloon catheter was advanced for dilation.Prior to procedure, the balloon did not track smoothly over a non-bsc guidewire, it was sticking.Then, upon inflation the balloon burst at 18 atmospheres.The device could not be removed, but the physician able to resheath through a 10f sheath and the procedure was completed.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18914106
Report Number2124215-2024-15671
Device Sequence Number6970920
Product Code FGE
UDI-Device Identifier08714729794868
UDI-Public08714729794868
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/27/2026
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0031133844
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/21/2024
Initial Report FDA Received Date03/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MERIT INQWIRE GUIDEWIRE
Patient SexUnknown
-
-