| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Following the reported event, the customer inspected the washer and found damage to the condensate return line allowing water to leak from the unit onto the floor.In addition, the customer identified that the washer's hose was damaged.The customer completed the repairs and returned the unit to service.A steris technician arrived onsite and confirmed the repairs have been made and the unit is operational.The dhr was reviewed for the unit and was found to be manufactured per specifications.A review of past complaints confirms this to be an isolated incident.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The user facility reported that water was leaking from their amsco 7052hp washer onto the floor.No report of injury.
|
| |
|