MAUDE Adverse Event Report: TELEFLEX MEDICAL WECK VISISTAT SKIN STAPLER 35R; STAPLE, REMOVABLE (SKIN)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL WECK VISISTAT SKIN STAPLER 35R; STAPLE, REMOVABLE (SKIN) Back to Search Results
Catalog Number 528135
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/16/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "hospital reported that our skin stapler got issues cannot work to close the wound".At the time of this report, the customer has not returned our requests for additional information.
 
Additional Manufacturer Narrative
(b)(4).The customer returned one unit of 528135 visistat 35r 6/box for investigation.Visual examination of the returned sample revealed that the staples appeared severely misaligned.Of these misaligned staples, four staples were observed to be pointed toward the proximal end of the stapler.The stapler was returned with 32 staples remaining in the cover block, indicating that at least 3 staples were fired by the customer.The trigger was returned partially engaged.No clear evidence of use in the form of biological material was present on the device.Dimensional inspection was not required as a part of this complaint investigation.Upon functional inspection, it appeared that the staple misalignment prevented the remaining staples from firing.The trigger cycle was completed, and no staples fired.Proper functional testing could not be performed due to the severe misalignment of the staples.The stapler was disassembled, and it was confirmed to have been assembled correctly.Die casting anomalies were discovered on the forming tool of the returned sample.No other defects or anomalies were observed.Per dhr the product visistat 35r 6/box lot # 73a2300516 was manufactured on 01/17/2023 a total of (b)(4) pieces.Lot was released on 02/01/2023.Dhr investigation did not show issues related to complaint.The ifu for this product was reviewed as a part of this complaint investigation.The ifu states, "to obtain optimum staple closure, the trigger must be squeezed all the way in." the sample was returned with its staples severely misaligned.Of these staples, four staples were observed to be pointed toward the proximal end of the stapler.Proper functional testing could not be performed due to the severe misalignment of the staples.The root cause of the staple misalignment could not be conclusively determined.A nonconformance was initiated to further investigate this complaint issue.The complaint of misfire/jamming - misaligned staples was confirmed based upon the sample received.Visual examination revealed that the staples appeared to be severely misaligned.Of these staples, two staples were observed to be pointed toward the proximal end of the stapler.Upon functional inspection, the misalignment of the staples prevented the remaining staples from firing correctly.Proper functional testing could not be performed due to the severe misalignment of the staples.The root cause of the staple misalignment could not be conclusively determined.Teleflex will continue to monitor and trend on complaints of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WECK VISISTAT SKIN STAPLER 35R
Common Device Name
STAPLE, REMOVABLE (SKIN)
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
MDR Report Key18914854
Report Number3003898360-2024-00298
Device Sequence Number6893023
Product Code QQS
UDI-Device Identifier14026704631770
UDI-Public14026704631770
Combination Product (Y/N)N
Initial Reporter CountryVM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number528135
Device Lot Number73A2300516
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2024
Initial Report FDA Received Date03/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-