| Model Number |
11400M25 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was learned via the patient registry that a patient with a 25mm 11400m mitris valve underwent a redo mitral valve replacement procedure after an implant duration of eight (8) days due to unknown reasons.The explanted valve was replaced with another 25mm 11400m mitris valve.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.An engineering evaluation is not required because there is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no use-related issue with a hazardous situation; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.An ifu review is unable to be performed, as no details regarding failure mode of the device were provided.A capa/scar/pra is not required as there are no confirmed product or labeling non-conformances and no other triggers are met.Based on the information available, a definitive root cause cannot be conclusively determined.
|
| |
|
Search Alerts/Recalls
|