| Model Number |
27040GP140-S |
| Medical Device Problem Codes |
Break (1069); Material Fragmentation (1261)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the bipolar cutting loop broke during procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Attempts were made to obtain additional information and to have device returned for evaluation; however, there has been no response from the reporter.Once the evaluation is completed or if new information is received, a supplemental report would be made to the fda.This complaint will be tracked and trended.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Changes made from the initial report: ---updated the udi number in section d4.Per investigation by the manufacturing site: according to the customer's description, it can be assumed that the cutting loop was pulled out in the distal area.The reason for this could be, for example, that the tensile force on the loop was too high without the hf current being activated or, on the other hand, that the activation time was too long and the wire burned out as a result.The ifu points out the correct handling.However, based on complaints in the past, it can be assumed that the user caused the fault.This event is filed under internal complaint id (b)(4).
|
| |
|