• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 6.5PLCF
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the tracheostomy tube's inflation line broke into pieces after it was pulled.There was no patient injury.
 
Additional Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.A visual inspection was performed, and it was observed the inflation line was missing the pilot balloon and was not received, the proximal end of the inflation line presented an even edge and a lack of bonding agent the failure mode pilot balloon separation was confirmed.It was reported that the tube's inflation line broke into pieces after it was pulled.The reported issue was confirmed.The most likely cause was determined to be manufacturing related.The instructions included with this device provide the following guidance: each tube¿s cuff, pilot balloon and valve should be tested by inflation prior to tube insertion.If disfunction is detected in any part of the inflation system, the tube should not be used.Do not overinflate the cuff.Ordinarily, cuff pressure should not exceed 25 cm of h2o.Closely monitor cuff pressure under a physician¿s guidance.Over inflation may lead to permanent tracheal damage, rupture of the cuff with subsequent deflation, or cuff distortion that may cause airway blockage.Avoid pulling or manipulation of the cuff inflation line as this could cause cuff deflation or removal of the tube from patient.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18917783
Report Number2936999-2024-00480
Device Sequence Number6760063
Product Code JOH
UDI-Device Identifier20884521779942
UDI-Public20884521779942
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K182861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6.5PLCF
Device Catalogue Number6.5PLCF
Device Lot Number23I0532JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2024
Supplement Date Received by Manufacturer07/21/2024
Initial Report FDA Received Date03/15/2024
Supplement Report FDA Received Date07/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-