|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.New information added to the following fields: b5, d9, h3, h4 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection, and the customer's reportable malfunction was confirmed.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the inability to grasp the tissue occurred due to the following mechanism: the tissue pad was worn out because the ultrasound was output with the gripper closed without gripping the tissue (including after the tissue was cut).Because the tissue pad was worn and peeled off, it was not possible to grasp the tissue well.The event can be prevented by following the instructions for use which state: do not activate output while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.When cutting or vessel sealing is performed, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.Olympus will continue to monitor field performance for this device.
|