| Model Number |
LTF-VP |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the image of the laparo-thoraco videoscope was blurry.The issue occurred during a procedure.There was no reported patient harm or impact due to this event.During device evaluation at olympus, it was found the image was foggy.The report is being submitted due to the defect found during evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and evaluated by olympus.In addition to the findings in b5, evaluation found the objective lens was leaking and there was fluid invasion, the objective lens adhesive was deteriorated, and the bending section cover adhesive was peeled.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although the foggy image was confirmed through evaluation of the device and was likely due to the objective lens leaking, a definitive root cause of the foggy image could not be determined.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the instructions for use (ifu).If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
|
| |
|
Search Alerts/Recalls
|