BIOMET UK LTD. OXFORD UNI PH3 TIB IMPACTOR; KNEE INSTRUMENT Back to Search Results
Catalog Number 32-420932
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a procedure, the surgeon broke the bolt of the tibial impactor.No piece fell into the patient.The surgery was completed with another device without any complications.This event is related to a malfunction that could potentially lead to a sterility issue/serious injury.However, no patient harm or further outcome was reported.Due diligence is in progress for this complaint; no further additional information or product has been received.
 
Additional Manufacturer Narrative
(b)(4) g2 - foreign: new zealand investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, d9, g3, g6, h2, h3, h4, h10.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.Visual examination of the returned product verified item and lot numbers matches the complaint.Scuffs and scratches noticed on the part.Dents and dings on strike plate.Bolt head is broken off the bolt.A review of the device manufacturing records confirmed no abnormalities or deviations.With the available information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXFORD UNI PH3 TIB IMPACTOR
Common Device Name
KNEE INSTRUMENT
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18920346
Report Number3002806535-2024-00096
Device Sequence Number18062879
Product Code HWR
UDI-Device Identifier05019279905638
UDI-Public(01)05019279905638(11)100917(10)ZB100801
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number32-420932
Device Lot NumberZB100801
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2024
Supplement Date Received by Manufacturer03/21/2024
05/17/2024
Initial Report FDA Received Date03/18/2024
Supplement Report FDA Received Date04/05/2024
05/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/17/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-