|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the subject device was manufactured.Based on the results of the investigation, the coating on the connecting tube was likely peeling do to wear/degradation from use over time.A root cause could not be determined.Instructions uretero-reno fiberscope olympus urf type p5 important information ¿ please read before use warnings and cautions.¿ do not strike, bend, hit, pull, twist, or drop the endoscope¿s distal end, insertion tube, bending section, and control section with excessive force.The endoscope may be damaged and could cause patient injury, such as burns, bleeding, and/or perforation.It could also cause parts of the endoscope to fall off inside the patient.¿ do not strike the distal end of the insertion tube or allow it to strike other objects.The objective lens surface at the distal end is particularly fragile, and visual abnormalities may result.¿ do not twist or bend the bending section with your hands.Equipment damage may result.¿ do not squeeze the bending section forcefully.The covering of the bending section may stretch or break and cause water leaks.Olympus will continue to monitor field performance for this device.
|