• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 11.5MM/8.5MM PROTECTION SLEEVE FOR RECON LOCKING; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 11.5MM/8.5MM PROTECTION SLEEVE FOR RECON LOCKING; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Catalog Number 03.010.075
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/27/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on an unknown date in 2024, the patient was treated with an frna gt nail.While implanting the nail, a 3.2 x 400 wire missed the hole through the nail.The wire was hitting the nail at an incorrect angle.Under x-ray imaging it was observed that the trajectory was the same as the tracer trajectory.This step of the surgical technique was repeated several times, creating an issue because the second recon screw wouldn¿t fit through the nail.This screw could not be implanted.No further information is available.This report is for an 11.5mm/8.5mm protection sleeve for recon locking.This is report 1 of 7 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: b3: event occurred on an unknown date in 2024.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: initial reporter is a synthes employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D10: therapy date is (b)(6), 2024.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was further reported that there wasn¿t a delay in surgery caused by this event.The frna implant has two different proximal locking options."recon locking¿ and the other ¿standard locking¿.Since recon locking wasn¿t working the surgeon was able to preform ¿standard locking¿.The surgery was completed successfully.There was no adverse patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11.5MM/8.5MM PROTECTION SLEEVE FOR RECON LOCKING
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18921895
Report Number8030965-2024-03743
Device Sequence Number7505627
Product Code FZX
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.010.075
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/28/2024
Supplement Date Received by Manufacturer04/16/2024
09/01/2026
Initial Report FDA Received Date03/18/2024
Supplement Report FDA Received Date04/18/2024
09/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
3.2MM TROCAR FOR RECON LOCKING; AIM-ARM RADIOLUC F/GREATER TROCHANTER FE; AIM-ARM RADIOLUC F/PIRIFORMIS FOSSA FEM; CANNULATED CONNECTING SCREW; DRILLSL 8.5/3.2 F/03.010.075; DRIVING CAP/THREADED; INSERT-HANDLE RADIOLUC FEM RECON NAIL
Patient Age18 YR
Patient SexMale
Patient Weight180 LB
-
-