• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. DEPTH DEV,2.7/3.5 /4.0MMLO-PRO; DEPTH GAUGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. DEPTH DEV,2.7/3.5 /4.0MMLO-PRO; DEPTH GAUGE Back to Search Results
Model Number DEPTH DEV,2.7/3.5 /4.0MMLO-PRO
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On (b)(6) 2024, it was reported by a sales representative via (b)(4) that an ar-8943-15 depth device, 2.7 / 3.5 / 4.0 mm, low profile from the titanium ankle fracture tray, ar-9943s, was not measuring correctly.Every screw was measured improperly because of the depth gauge.The gauge is supposed to be 20mm according to what it is reading in the back.It is 2mm off.This caused every screw to be off.This was discovered during a distal fibula fx with syndesmosis and deltoid repair with no patient harm.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h6.Complaint allegation is not confirmed.Dimensional testing: components were measured and met dimension.690 ±.010 per drawing c6043 rev 11.Visual inspection revealed abrasion marks on the tip of the depth device, 2.7 / 3.5 / 4.0 mm, low profile.Functional testing was performed, and no issues were found with the slider pin that translate smoothly into the body.No problem found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH DEV,2.7/3.5 /4.0MMLO-PRO
Common Device Name
DEPTH GAUGE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key18922720
Report Number1220246-2024-01524
Device Sequence Number15055388
Product Code HTJ
UDI-Device Identifier00888867052253
UDI-Public00888867052253
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 07/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDEPTH DEV,2.7/3.5 /4.0MMLO-PRO
Device Catalogue NumberAR-8943-15
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2024
Supplement Date Received by Manufacturer06/12/2024
Initial Report FDA Received Date03/18/2024
Supplement Report FDA Received Date07/01/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-