| Model Number |
DEPTH DEV,2.7/3.5 /4.0MMLO-PRO |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, it was reported by a sales representative via (b)(4) that an ar-8943-15 depth device, 2.7 / 3.5 / 4.0 mm, low profile from the titanium ankle fracture tray, ar-9943s, was not measuring correctly.Every screw was measured improperly because of the depth gauge.The gauge is supposed to be 20mm according to what it is reading in the back.It is 2mm off.This caused every screw to be off.This was discovered during a distal fibula fx with syndesmosis and deltoid repair with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h6.Complaint allegation is not confirmed.Dimensional testing: components were measured and met dimension.690 ±.010 per drawing c6043 rev 11.Visual inspection revealed abrasion marks on the tip of the depth device, 2.7 / 3.5 / 4.0 mm, low profile.Functional testing was performed, and no issues were found with the slider pin that translate smoothly into the body.No problem found.
|
| |
|