| Model Number |
DISPOSABLE KIT, KNEE FIBERTAKS |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 02/26/2024, it was reported by a sales representative via (b)(4) that an ar-3710 disposable fibertak anchor pulled out.This occurred during use in a case with no patient effect.Both anchors came out of the bone while the surgeon was tensioning the loop suture to the ligament.A separate disposable kit was utilized for each anchor due to a drop on the sterile field.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause can be attributed to misaligned insertion; or prying/leveraging the device during use.
|
| |
|