| Catalog Number |
0620040610 |
| Medical Device Problem Codes |
No Flow (2991); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of insufflation.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: frozen screen error no menu backlight is out blank screen.Probable root cause: 1.Software malfunction.2.Power supply malfunction.3.Insufflator design.4.Unwanted movement of internal components / wiring.5.Use error.6.Electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge.7.Mio board malfunction.8.Bam board malfunction.9.Lcd assembly failure.10.Manufacturing/ service error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of insufflation.
|
| |
|
Search Alerts/Recalls
|