• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR; INSUFFLATOR, LAPAROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR; INSUFFLATOR, LAPAROSCOPIC Back to Search Results
Catalog Number 0620040610
Medical Device Problem Codes No Flow (2991); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was loss of insufflation.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: frozen screen error no menu backlight is out blank screen.Probable root cause: 1.Software malfunction.2.Power supply malfunction.3.Insufflator design.4.Unwanted movement of internal components / wiring.5.Use error.6.Electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge.7.Mio board malfunction.8.Bam board malfunction.9.Lcd assembly failure.10.Manufacturing/ service error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Event or Problem Description
It was reported that there was loss of insufflation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR
Common Device Name
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key18923327
Report Number0002936485-2024-00236
Device Sequence Number12638746
Product Code HIF
UDI-Device Identifier07613327063295
UDI-Public07613327063295
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K063367
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number0620040610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/28/2024
Supplement Date Received by Manufacturer02/28/2024
Initial Report FDA Received Date03/18/2024
Supplement Report FDA Received Date06/20/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-