Model Number E10503
Medical Device Problem Codes Smoking (1585); Arcing (2583); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Burn(s) (1757); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during procedure, the generator had a small smoke and spark incident that caused burn.The spark occurred at the portion of the monopolar cord hooked up with the non medtronic lap instrument.They ran pm checks on the generator and it passed.The team in the room believed that someone was activating the foot pedal for an extended period of time that caused the spark.
 
Additional Manufacturer Narrative
D10 concomitant product: vlft10gen, vlft10gen ft series energy platformx1, (sn: (b)(6)).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
This event has been reassessed and found not to be a reportable event and is not associated with a serious injury or potential for serious injury with reoccurrence.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during procedure, the monopolar cord had a small smoke and spark incident that caused burn to the cord.The extracorporeal spark occurred at the portion of the monopolar cord hooked up with the non medtronic lap instrument.They ran pm checks on the generator and it passed.The team in the room believes that some one was activating the foot pedal for an extended period of time that caused the spark.They used a new cord with same generator and it worked fine.There was no burn direct the patient.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18924307
Report Number1717344-2024-00727
Device Sequence Number20366026
Product Code GEI
UDI-Device Identifier10884521782631
UDI-Public10884521782631
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K193004
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE10503
Device Catalogue NumberE10503
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2024
Supplement Date Received by Manufacturer03/29/2024
Initial Report FDA Received Date03/18/2024
Supplement Report FDA Received Date04/24/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-