• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 2.0MM DRILL BIT/QC/100MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.0MM DRILL BIT/QC/100MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 310.19
Medical Device Problem Codes Break (1069); Entrapment of Device (1212); Activation, Positioning or Separation Problem (2906); Material Integrity Problem (2978)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Insufficient Information (4580)
Date of Event 02/29/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d1, d2, d3, d4, g4-510k: this report is for an unknown drill bits/unknown lot.Part and lot number are unknown.Without the specific part number; the udi number and 510-k number is unknown.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, a broken drill left in the patients hip.This report is for one (1) unknown drill bits.This is report 1 of 1 for complaint (b)(4).
 
Event or Problem Description
The surgery was completed successfully with no delay.There is no plan for a revision surgery.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A1-a4: patient information added.B3: event date added.B5: event description updated.D4: udi, lot number, catalog code added.G4: 510k: exempt.H4: manufacture date added.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D1: brand name updated.D2a, d2b: procode, common device name updated h6: medical device problem codes updated.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: part number: 310.19-us, lot number: 3455p74, part manufacture date: 06-dec-2022, manufacturing location: elmira, part expiration date: n/a, nonconformance noted: n/a.Dhr record review: a review of the device history record revealed no complaint related anomalies.The device history record shows this lot of 2.0mm drill bit/qc/100mm was processed through the normal manufacturing and inspection operations in elmira with no rework nor nonconformities noted.This lot met all dimensional, visual, and packaging criteria at the time of release from elmira with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed this lot met all specifications with no nonconformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.This lot met all dimensional, visual, and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the shaft of the 2.0mm drill bit/qc/100mm was broken, fragment was not received for evaluation.No x-ray evidence was received in order to confirm the embedded device allegation.No other issues were identified.A dimensional inspection for the 2.0mm drill bit/qc/100mm was not performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the 2.0mm drill bit/qc/100mm would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM DRILL BIT/QC/100MM
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
ELMIRA
35 airport road
horseheads NY 14845
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18924597
Report Number8030965-2024-03785
Device Sequence Number19952999
Product Code HWE
UDI-Device Identifier10886982186102
UDI-Public(01)10886982186102
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number310.19
Device Lot Number3455P74
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/01/2024
Supplement Date Received by Manufacturer03/28/2024
04/30/2024
Initial Report FDA Received Date03/18/2024
Supplement Report FDA Received Date04/16/2024
04/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/09/2022
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient Age51 YR
Patient SexMale
Patient Weight113 KG
-
-