It was reported that, during total ankle replacement surgery, the surgeon broke off a cadence implant protector - right protector while removing from implant.It is unknown if pieces fell into the patient, but surgery was resumed without any delay.No patient injuries were reported.
|
|
H3, h6: the associated device was returned and evaluated.The visual inspection revealed that the reusable instrument had a piece fractured off.Visual evaluation identified signs of wear due to normal use and scratches along the device.Notably, there were some indentations on the top of the protector near the u-shaped opening on opposing sides.On the back side, the edge seemed to be dented/deformed.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed a similar event for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|