ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY VR, U1.6 DF1 US; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number CD1231-40
Medical Device Problem Codes Premature Discharge of Battery (1057); Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient's implantable cardioverter defibrillator (icd) exhibited inappropriate shock due to incorrect interpretation of signal during atrial fibrillation episodes.Premature battery depletion was also noted.The device was explanted and replaced.The patient was stable.
 
Additional Manufacturer Narrative
The reported field event of inappropriate therapy, sensing issue and premature battery depletion was not confirmed.As received, the device was at end of service (eos) upon receipt.Review of the device data showed that the device delivered many consecutive hv therapies in a short period of time.The hv therapies were delivered appropriately, and the device acted normally given its programmed settings.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Telemetry, impedance, sensing, pacing, capacitor maintenance, and high voltage (hv) output functions of the device were tested and found to be normal.No anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY VR, U1.6 DF1 US
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18926949
Report Number2017865-2024-35285
Device Sequence Number19981341
Product Code NVZ
UDI-Device Identifier05414734504355
UDI-Public05414734504355
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2021
Device Model NumberCD1231-40
Device Lot NumberS000074202
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer03/20/2024
04/11/2024
04/25/2024
Initial Report FDA Received Date03/18/2024
Supplement Report FDA Received Date03/21/2024
04/25/2024
04/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/26/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-