• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG SL-PLUS MIA CURVED RASP; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG SL-PLUS MIA CURVED RASP; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 75004495
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during a thr surgery, one (1) sl-plus mia curved rasp broke.The procedure was resumed, without any delay, using a s+n back-up device.No injury was reported as a consequence of this issue.
 
Additional Manufacturer Narrative
H3, h6: it was reported that, during a total hip replacement surgery, one (1) sl-plus mia curved rasp broke.No injury was reported as a consequence of this issue.The complaint device has not been completely returned.A visual evaluation of the returned part was conducted, and it was concluded that there is a fracture at the distal part of the device.The distal tip fractured off and is missing.Furthermore, there are signs of wear such as scratches and dents.A review of the production documentation did not detect any deviation that could have contributed to the reported failure mode.The complaint history review for the complaint device revealed no additional complaints for the affected batch nor additional complaints for the same product number over the past 12 months with similar failure mode.A review of past corrective actions was performed.No further escalation is required.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.There is no indication that the reported device failed to meet manufacturing specifications upon release for distribution.The root cause of the reported event remains undetermined.Considering the performed visual inspection, it is suspected that the mentioned device failure arose from a ¿wear and tear" issue.Normal wear and tear are known to contribute to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag", all devices must be inspected and controlled for proper functioning after cleaning/disinfection.The need for further actions is not indicated.Nevertheless, smith+nephew will continue to monitor this device for similar issues.The returned device will be discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL-PLUS MIA CURVED RASP
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18927177
Report Number9613369-2024-00026
Device Sequence Number11907218
Product Code JDI
UDI-Device Identifier07611996081152
UDI-Public07611996081152
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number75004495
Device Lot Number21000138
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer04/09/2024
Initial Report FDA Received Date03/18/2024
Supplement Report FDA Received Date04/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-