Catalog Number 130-32-53
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Cyst(s) (1800); Failure of Implant (1924); Pain (1994); Osteolysis (2377); Ambulation or Postural Difficulties (2544)
Date of Event 02/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H3: pending investigation.D10: 4179599 180-01-56 - crown cup, cluster-hole.H7: z-2117-2021.
 
Event or Problem Description
As part of the manufacturer's recall campaign, the patient presented for a new check-up.In the x-ray and ct control showed the well-known clear decentering of the prosthetic head and a clear increase in size of the known lysis in the acetabulum.That's why the revision and change took place on (b)(6) 2024 on a vite-hardened, specially approved inlay (novation xle neutral liner, group 3, 36 mm i.D., ref 140-36-53, sn: (b)(6).As part of the replacement operation, in addition to the inlay replacement, the firmness was determined cup integrity as well as the curettage and sealing of the cysts in the socket using allogeneic cancellous bone and changing the prosthetic head.
 
Additional Manufacturer Narrative
H3: a number of variables including use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential) could have all contributed to the increased trend in wear/osteolysis related complaints.The most likely cause for the revision reported due to early prosthesis wear/osteolysis is a combination of the risk factors specified in the hhe.However, this cannot be confirmed from the reported information.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The most likely cause for the revision reported due to early prosthesis wear/osteolysis is a combination of the risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed from the reported information.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: e.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b5, d4.The following sections were corrected: h6.
 
Event or Problem Description
It was reported via ous legal documentation that a patient underwent a total hip replacement.During the subsequent assessments and examinations conducted by the hospital it was determined that the prosthetic head required revision.It became apparent that the plastic inlay had already suffered significant wear and tear, and there was a risk that the plastic particles from the dissolving implant could damage the surrounding bone.For this reason, our client had to undergo a maximally invasive revision surgery, followed by several weeks of postoperative treatment.Intraoperatively, premature wear of the inlay was confirmed.Since the replacement of the defective hip prosthesis and the bone damage caused by the material, the patient has experienced severe pain during prolonged normal activity, such as walking.As a result, normal movement has been severely limited since then.This, in turn, has had a huge impact on the performance of daily tasks and social activities and has significantly reduced the patient's quality of life.The patient has suffered from severe pain and restricted mobility.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6, h11.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/ technique.A definitive root cause was unable to be determined; however, may have resulted from use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LNR, NEUTRAL, 32MM ID, GROUP 3 CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
michael crader
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key18927394
Report Number1038671-2024-00547
Device Sequence Number14581129
Product Code JDI
UDI-Device Identifier10885862022172
UDI-Public10885862022172
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2020
Device Catalogue Number130-32-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2024
Supplement Date Received by Manufacturer05/22/2024
12/13/2024
05/14/2025
02/06/2025
08/29/2025
Initial Report FDA Received Date03/18/2024
Supplement Report FDA Received Date05/29/2024
12/16/2024
05/14/2025
08/15/2025
09/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/03/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-