| Model Number |
CC60WF |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that during an intraocular lens (iol) implant procedure, the lens was loaded and injected into the patient eye.It was observed that lens on the left side was split by the haptic also there was a crack on the right side in two areas.The lens was explanted in the initial procedure.Additional information has been received new lens was implanted and there was no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The account indicated the use of qualified associated products.The product investigation could not identify a root cause for the reported complaint.The product has not been received to evaluate.The account indicated the use of qualified associated products.The instructions for use (ifu) instructs: the ifu instructs to completely fill the cartridge with ophthalmic viscosurgical device (ovd) immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.No further information has been provided at this time.File will be reopened when new information or the product is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|