MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PAEDIGAV 4/24; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PAEDIGAV 4/24; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV298T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 02/29/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation: visual inspection: during the investigation, no significant deformation or damage of the valve was determined.Permeability test: a permeability test has shown that the valve is permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve does not operate within the accepted tolerance in either position.An accelerated outflow of paedigav could be determined.Internal inspection: after dismantling of the valve, deposits were found.Results: based on our investigation results, we can determine an accelerated outflow in the valve.The determined deposits can be named as the cause for the accelerated outflow.Organic material in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.
 
Event or Problem Description
It was reported that a paedigav (#(b)(6)) was implanted during a procedure performed in 2019.According to the complainant, the valve caused an underdrainage.The patient underwent a revision procedure.The complainant device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 6 years, height: 108 cm, weight: 19 kg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAEDIGAV 4/24
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key18928780
Report Number3004721439-2024-00073
Device Sequence Number17278346
Product Code JXG
UDI-Device Identifier04041906116330
UDI-Public4041906116330
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K011030
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/21/2022
Device Model NumberFV298T
Device Catalogue NumberFV298T
Device Lot Number20035113
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2024
Initial Report FDA Received Date03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/19/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age6 YR
Patient SexUnknown
Patient Weight19 KG
-
-