The initial malfunction occurred on the artis zee floor unit on march 6, 2024.The on-site service interventions following the incident showed a ¿potential write performance issue¿ indicating an problem with the image disk and/or the copra board.Therefore, the image disk and the copra board were replaced as part of service activity.However, the issue re-occurred on may 15, 2024, (mdr # 3004977335-2024-00059) during a heart replacement procedure.There were no health consequences associated with this event.Further detailed investigation showed sporadic issues with the sas (serial attached scsi) controller.The sas controller serves as the interface between storage devices and the computer's motherboard, facilitating data transfer between them.The log-file showed that when the x-ray was performed, the image file system component in the image acquisition system (ias) was not able to write x-ray data on the image drive due to a timeout failure.Furthermore, system event log showed a sas controller error for the image drive.Subsequent, the system was not able to produce x-ray.To prevent any further recurrences, the entire image acquisition system pc was replaced on may 23,2024.However, the investigation did not identify any hardware problems.The system was kept under observation for approximately 2 months.The error has not reoccurred.Clear root cause of the issue could not be determined.According to the experts, a contact problem was the most likely cause of the observed behavior, which may have been remedied by vibrations during transport.The primary udi number in section d4 was corrected.
|