| Catalog Number |
600-9005-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On 08-mar-2024, abbott point of care (apoc) was contacted by a customer regarding i-stat chem8+ cartridge that yielded a discrepant potassium result on a patient.There was no additional patient information at the time of this report.Return product is available for investigation.Method date tested result(mmol/l) sample i-stat (b)(6) 2024 23:00 6.0, a; i-stat (b)(6) 2024 ni 7.7, a; catalyst (b)(6) 2024 ni >10, a.Customer reported that the patient at this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggested that product was not performing within the variability.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 15-apr-2024.A review of the device history record (dhr) confirmed the cartridge lot met finished goods release criteria.Retained cartridge testing met the acceptance criteria found in q04.01.003 rev.An, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been determined for chem8+ cartridge lot h23287.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|