The following devices were also listed in this report: exeter v40 stem 44mm no1l125; cat# 0580-3-441; lot# g8211609 26mm +12 lfit v40 head; cat# 6260-9-426; lot# 82320103 it cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Reported event: an event regarding dislocation & infection involving a uhr head was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock on with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot or sterile lot referenced.Conclusion: it was reported that the patient was revised due to dislocation, incorrect selection & infection.All stryker products sold as sterile are validated to a minimum sterility assurance level sal of 10^-6 in accordance with applicable iso standards.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|