| Model Number |
A22255C |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
It was reported the tip of the high frequency resection electrode broke while inside the patient during a hysteroscopy/dilation and curettage (d&c) procedure.The broken portion was retrieved.The procedure was completed with the same device.The device was discarded.There was no reported patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.Since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.However, the probable cause of the malfunction was due to wear and tear / the loop at the distal end of the electrode wears out during use and can break, burn or melt.The lot/serial number for the affected device could not be obtained.Therefore, a dhr review could not be performed.Olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide a correction to the fields (h6) and to provide a correction with information inadvertently left out (g2 and h6).Correction to the previous h6 as 3331.And 10 were incorrectly selected.
|
| |
|