| Model Number |
02-00139 |
| Medical Device Problem Code |
Biocompatibility (2886)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Pain (1994); Rash (2033); Skin Inflammation/ Irritation (4545)
|
| Date of Event |
02/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, the patient was experiencing a rash while wearling the lead wire adapater (lwa).The patient noted that the the rash had spread to the nipple/breast area and bruising was present.They were experiencing bleeding, scrappiness, oozing and rawness that is very painful.The patient contacted their dermatologist regarding the rash and they were prescribed doxycycline to take to relieve irritation.The patient's cardiologist advised the patient to continue monitoring.The patient reported that they have numerous allergies to triple ointment ingredients and latex and a mild adhesive allergy.It is thought the skin irritation is due to the electrodes.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the patient experienced severe skin irritation while wearing the lead wire adapter.The device was not returned for investigation.Engineering evaluation was unable to be performed on the electrode as it is single use and was not returned.Allegation is confirmed prescription from a medical professional and is most probable to be a bio-incompatibility issue with the electrode adhesive and/or hydrogel components.Marsi, skin burn, and associated symptoms may inherently occur under the course of ecg monitoring.No single factor or combination of factors can be attributable to electrode skin irritation and associated symptoms.The product labeling advises patients of alternate options and other steps to take if skin irritation develops, including healthcare professional contact as needed.
|
| |
|
Search Alerts/Recalls
|