TELEFLEX MEDICAL DEK BL MF 0 TC-43/HR26 2N 36" BSC; DEKLENE MAXX POLYPROPELENE NON Back to Search Results
Catalog Number 833-235
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Customer complaint cannot be confirmed based only on the information provided, to perform a proper evaluation and determine the source of defect reported it is necessary to evaluate the sample involved on this complaint.If the sample becomes available this investigation will be updated with the evaluation results.A device history record review was performed, and no relevant findings were identified.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that "capio went into ssl the bullet stayed in the ligament and came off the suture.The bullet could not be palpated because it was deep in the ligament.Location was confirmed by intraoperative x-ray.".No patient harm or health consequences.The current status of the patient is reported as "no health consequences".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEK BL MF 0 TC-43/HR26 2N 36" BSC
Common Device Name
DEKLENE MAXX POLYPROPELENE NON
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
keona jones
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18935640
Report Number3004365956-2024-00013
Device Sequence Number15175250
Product Code GAW
UDI-Device Identifier14026704735324
UDI-Public14026704735324
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K153076
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number833-235
Device Lot Number74C2202885
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2024
Initial Report FDA Received Date03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED
Patient SexUnknown
-
-