| Model Number |
UNI GLENOID-PERIPHERAL LOCK SCRW |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during an eclipse with universal glenoid surgery during implantation of the universal glenoid baseplate the inferior peripheral screw got jammed.When trying to screw it in, the head of the screwdriver broke off and got stuck in the screw head.This had to be removed using the operace system.The screw head broke off in the process.The rest of the screw remained in the patient and the baseplate was left as it was.The inlay was then placed on top and the surgery was completed as normal.There was no harm for patient, operator or third party.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to over-torquing/over-engaging the inserter within the screw during use.
|
| |
|