• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDLINE EXCELSIOR; SYRINGE, 10ML SALINE IN 10ML

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDLINE EXCELSIOR; SYRINGE, 10ML SALINE IN 10ML Back to Search Results
Model Number EMZE010001H
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that the cap of the saline flush syringe was difficult to remove.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.Evaluation of the reported problem/issue was performed using a returned sample and retained samples.Tip cap torque testing was performed on all samples and all syringes were found to be within specification.In an abundance of caution, and in response to an fda 483 issued for cfn 1417592 on 22-jan-2024, this medwatch is being filed.If additional relevant information becomes available a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported that the cap of the saline flush syringe was difficult to remove.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SYRINGE, 10ML SALINE IN 10ML
Manufacturer (Section D)
MEDLINE EXCELSIOR
1933 heck drive
neptune NJ 07753 4428
Manufacturer Contact
nigel vilches
three lakes drive
northfield, IL 60093-2753
MDR Report Key18937350
Report Number2027791-2024-00038
Device Sequence Number6971104
Product Code NGT
UDI-Device Identifier00363807100015
UDI-Public00363807100015
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberEMZE010001H
Device Lot Number3133787
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/24/2019
Initial Report FDA Received Date03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-