• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXTRACTOR PRO RX; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXTRACTOR PRO RX; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00547020
Medical Device Problem Codes Difficult to Advance (2920); Adverse Event Without Identified Device or Use Problem (2993); Failure to Deflate (4060)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/27/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a140101 captures the reportable event of balloon failure to deflate.
 
Event or Problem Description
It was reported to boston scientific corporation that an extractor pro rx retrieval balloon was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.During the procedure, there was very strong resistance when the balloon was inserted into the bile duct.The balloon could be inflated without any problem, but could not be deflated, so it was removed from the bile duct by force.The procedure was completed with another extractor pro rx retrieval balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a140101 captures the reportable event of balloon failure to deflate.Block h10: investigation results.The returned extractor pro rx retrieval balloon was analyzed, and a microscopic and functional inspection found the balloon was not capable of being inflated and deflated due to it was occluded most likely with dry contrast media, to remove the dry contrast media the device was submerged in warm water, and then the device was inspected again but the problem remained.A wire was introduced to see where the device was occluded and then cutting specifically in that location and after the destructive inspection, it was observed that the catheter has a lump of contrast media restricting the capability of the device to have its functionality tested.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon failure to deflate was unable to be confirmed.The results of the analysis performed on the returned device found some dry contrast media inside the catheter and the balloon, restricting the capability of the device to have its functionality tested.Therefore, the most probable root cause is cause not established.
 
Event or Problem Description
It was reported to boston scientific corporation that an extractor pro rx retrieval balloon was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.During the procedure, there was very strong resistance when the balloon was inserted into the bile duct.The balloon could be inflated without any problem, but could not be deflated, so it was removed from the bile duct by force.The procedure was completed with another extractor pro rx retrieval balloon.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACTOR PRO RX
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18937732
Report Number3005099803-2024-01073
Device Sequence Number15196315
Product Code FGE
UDI-Device Identifier08714729790310
UDI-Public08714729790310
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K102082
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model NumberM00547020
Device Catalogue Number4702
Device Lot Number0030964298
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer04/23/2024
Initial Report FDA Received Date03/19/2024
Supplement Report FDA Received Date05/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-