• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORPORATION 1.8MM Q-FIX DISPOSABLE FLEXIBLE DRILL; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORPORATION 1.8MM Q-FIX DISPOSABLE FLEXIBLE DRILL; TEMPLATE Back to Search Results
Catalog Number 72290118
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during a labral repair, the q-fix drill was physically broken.The procedure was completed using a back-up device.There was a surgical delay less than 30 minutes and no further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4) h10 h3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.8MM Q-FIX DISPOSABLE FLEXIBLE DRILL
Common Device Name
TEMPLATE
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18937811
Report Number3006524618-2024-00111
Device Sequence Number6971588
Product Code HWT
UDI-Device Identifier00885556657447
UDI-Public00885556657447
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/26/2026
Device Catalogue Number72290118
Device Lot Number2112295
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2024
Supplement Date Received by Manufacturer04/09/2024
Initial Report FDA Received Date03/19/2024
Supplement Report FDA Received Date04/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/26/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-