Catalog Number MSWSTDA35180
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect medical device is not expected to return for evaluation.The complaint could not be confirmed.The root cause could not be determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.Should the device be returned later, the investigation will be reopened.
 
Event or Problem Description
The account alleges that during a procedure, the clear hydrophilic coating was coming off the guidewire after activating the device for procedural use.No patient contact or injury to report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPLASHWIRE¿.GUIDE WIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
MERIT MEDICAL SYSTEM IRELAND
parkmore business park west
ballybrit,
galway
EI 
Manufacturer (Section G)
MERIT MEDICAL SYSTEM IRELAND
parkmore business park west
ballybrit,
galway
EI  
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
8012088662
MDR Report Key18938071
Report Number9616662-2024-00014
Device Sequence Number6886725
Product Code DQX
UDI-Device Identifier00884450529669
UDI-Public884450529669
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K201595
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMSWSTDA35180
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2024
Initial Report FDA Received Date03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-