| Catalog Number |
00595109000 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign: germany.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Upon receipt of additional information or completion of the investigation, a follow-up mdr will be submitted.H3 other text : see h10 narrative.
|
| |
|
Event or Problem Description
|
|
It was reported that during an initial total knee arthroplasty, the tibial broach impacting instrument fractured at the weld point.There was no patient impact and no adverse events have been reported as a result of the malfunction.Due diligence is in progress for this event; to date no further information has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections have been updated.H6: proposed component code: mechanical (g04) - impactor visual evaluation of the provided photos found signs of repeated use (gouges).The weld that connects the subcomponents has fractured.The instrument has an approximate field age of 7+ years.Device history record was reviewed and no discrepancies related to the reported event were found.Investigation results concluded that the reported event is attributed to the seven plus (7+) years of use in the field and the normal wear associated with it.Complaint sample was evaluated and the reported event was confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; d9; g3; h2; h3; h6; h11.H6: proposed component code: mechanical (g04) - impactor visual evaluation of the returned product/provided photos found signs of repeated use (gouges, tool marks).Evaluation confirms that the weld connecting the subcomponents has fractured.All pieces have been returned.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.Device history record was reviewed and no discrepancies related to the reported event were found.Investigation results concluded that the reported event is attributed to the seven plus (7+) years of use in the field and the normal wear and tear associated with repeated use over time.Complaint sample was evaluated and the reported event was confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.
|
| |
|